Standard Operating Procedures
BIOSIGN SSI Test Pack
Last updated: January 5, 2012
I. Purpose
The Getinge BIOSIGN SSI Test Pack is specifically designed for biological testing and monitoring of 132oC pre-vacuum steam sterilization cycles. The product gives the user instant reassurance based on a result from a steam integrator card, and after biological indicator incubation, even greater assurance that the sterilizer operated at proper sterilization parameters.
II. General Information
- Storage: Store test packs in a refrigerator ranging from 2-24° C.
- Frequency of use: once monthly.
III. BIOSIGN SSI Test Pack
- BIOSIGN Instant Read Integrator Record Card.
- Provides immediate verification that the test pack was exposed to correct sterilization parameters.
- Ink color on card changes from PURPLE (FAIL) to GREEN (PASS)
- Instant results
- BIOSIGN Biological Indicator.
- Includes a paper strip containing G. stearothermophilus inside an ampule of specially formulated growth medium.
- Sterilization failure is noted by positive growth of the test strip turning the medium color from red to yellow.
- Results in 24 hours. YELLOW (FAIL), RED (PASS).
- Place one Test Pack in autoclave
- Place window side up on the lowest shelf and above the drain.
- Do not set any thing on top of Test Pack.
- Process the load using the "bedding" cycle.
IV. Procedure for Using the BIOSIGN SSI Test Pack
- Turn Biosign incubator on 45 minutes before needed in order to reach 55oC. Use caution as the metal heating block will be hot.
- Remove 2 Test Packs box from refrigerator in HS009.
- One will be autoclaved, the other kept as a control.
- Medium in test pack vial should be RED.
- Place 1 Test Pack box in autoclave.
- Place window side up on the lowest shelf and above the drain.
- Do not set anything on top of Test Pack.
- Process the load using the "bedding" cycle.
- Remove Test Pack box from autoclave at end of cycle.
- Allow box to cool for at least 15 minutes.
- Remove and examine the Integrator Record Card. An ink color change on the card from PURPLE to GREEN indicates correct exposure conditions of temperature, time and steam.
- Save the card, label with date and file with other sterilization processing records.
- Remove the Biological Indicator vial from the box and process as directed below.
- Activate the Biological Indicator and incubate in Biosign Incubator (see below).
- Be certain the autoclaved vial has cooled to room temperature.
- Insert both autoclaved and control vials into 55oC incubator.
- Insert (at an angle) the bottom (only) of vial into an empty well in the incubator.
- Ensure the bottom of the vial contacts the bottom of the well.
- Pivot (snap up) the vial into the well such that the edges of the well crush the vial.
- Close the incubator cover.
- Incubate vials for 24 hours.
- The medium in the autoclaved vial should be RED (dead spores).
- The medium in the control vial should be YELLOW (live spores).
- Record findings in BIOSIGN record book in HSB 016.
- React to positive growth (YELLOW) in the autoclaved vial by having the autoclave checked immediately.
- Before disposal, autoclave all the vials for at least 30 minutes on the "Bedding" setting.