I. Purpose

The Getinge BIOSIGN SSI Test Pack is specifically designed for biological testing and monitoring of 132oC pre-vacuum steam sterilization cycles.  The product gives the user instant reassurance based on a result from a steam integrator card, and after biological indicator incubation, even greater assurance that the sterilizer operated at proper sterilization parameters.

II. General Information

  • Storage: Store test packs in a refrigerator ranging from 2-24° C.
  • Frequency of use: once monthly.

III. BIOSIGN SSI Test Pack

  • BIOSIGN Instant Read Integrator Record Card.
    • Provides immediate verification that the test pack was exposed to correct sterilization parameters.
    • Ink color on card changes from PURPLE (FAIL) to GREEN (PASS)
    • Instant results 
  • BIOSIGN Biological Indicator.
    • Includes a paper strip containing G. stearothermophilus inside an ampule of specially formulated growth medium.
    • Sterilization failure is noted by positive growth of the test strip turning the medium color from red to yellow.
    • Results in 24 hours.  YELLOW (FAIL), RED (PASS).
  • Place one Test Pack in autoclave
    • Place window side up on the lowest shelf and above the drain.
    • Do not set any thing on top of Test Pack.
    • Process the load using the "bedding" cycle.

IV.  Procedure for Using the BIOSIGN SSI Test Pack

  • Turn Biosign incubator on 45 minutes before needed in order to reach 55oC.  Use caution as the metal heating block will be hot.
  • Remove 2 Test Packs box from refrigerator in HS009.
    • One will be autoclaved, the other kept as a control.
    • Medium in test pack vial should be RED.
  • Place 1 Test Pack box in autoclave.
    • Place window side up on the lowest shelf and above the drain.
    • Do not set anything on top of Test Pack.
    • Process the load using the "bedding" cycle.
  • Remove Test Pack box from autoclave at end of cycle.
    • Allow box to cool for at least 15 minutes.
    • Remove and examine the Integrator Record Card.  An ink color change on the card from PURPLE to GREEN indicates correct exposure conditions of temperature, time and steam.
      • Save the card, label with date and file with other sterilization processing records.
    • Remove the Biological Indicator vial from the box and process as directed below.
  • Activate the Biological Indicator and incubate in Biosign Incubator (see below).
    • Be certain the autoclaved vial has cooled to room temperature.
    • Insert both autoclaved and control vials into 55oC incubator.
      • Insert (at an angle) the bottom (only) of vial into an empty well in the incubator.
      • Ensure the bottom of the vial contacts the bottom of the well.
    • Pivot (snap up) the vial into the well such that the edges of the well crush the vial.
    • Close the incubator cover.
    • Incubate vials for 24 hours.
      • The medium in the autoclaved vial should be RED (dead spores).
      • The medium in the control vial should be YELLOW (live spores). 
    • Record findings in BIOSIGN record book in HSB 016.
    • React to positive growth (YELLOW) in the autoclaved vial by having the autoclave checked immediately.
    • Before disposal, autoclave all the vials for at least 30 minutes on the "Bedding" setting.